Please Register and Log In !   

Novaspec® III+ Spectrophotometer

Large numerical display
Absorbance, concentration & kinetics modes
Additional cell holder accessories
Login/Register to shop in US only.
All others please contact us for local pricing/quotes.

 

Novaspec® III+ Spectrophotometer

Basic enzyme kinetics - available now!

Product Code Price
Novaspec III+ 80-2120-40      View Price
Novaspec III+ Ext Warranty 80-5000-01      View Price
Novaspec Pro / Novaspec III+ Service      View Price

 

Click here for details.
OR

Details

Novaspec III+ has been designed for use in the student teaching laboratory. It is a rugged lightweight instrument with no moving parts, making it both portable and highly reliable. Automatic calibrations at startup gives confidence in its use.

It measures absorbance, % transmission and concentration as well as simple kinetics. The large display allows the user to view results in numerical value as well as a graphical view. The user can view the peak, zoom in and out all on the unit.

Up to 72 methods can be stored providing the user a quick and easy way to recall frequently used methods. These methods can also be locked using a pin protected password, as well as unlocked and deleted.

The Novaspec III+ also provides the ability to label your sample for easy identification. Your sample can be measured in a test tube using the optional test tube cell holder accommodating test tubes that are 10 to 18 mm in diameter. A heated cell holder is also available for thermostatted measurements at 37°C.

PVC (Print Via Computer) software is included with the instrument providing the user with the option to transfer data directly to a PC. Data can be easily stored on a USB memory stick and analyzed at a later date

Single beam visible spectrophotometer covering 325 - 1100 nm
Save data on a USB memory stick and analyze results later
Test tube and heated cell holder accessories are available
PVC (Print Via Computer) software included to allow data transfer to PC from the instrument
Light Source Pulsed Tungsten Halogen
Detection Technology CMOS Array
Spectral Bandwidth <7 nm
Photometric Reproducibility < 0.002 A at 0 A and 500 nm
Photometric Min. -0.300 A
Stability ± 0.005 A/h near 0 A
Stray Light < 1.0%T at 340 nm (ANSI/ASTM E387-72)
Wavelength 325-1100 nm
Accuracy ± 2 nm
Absorbance -0.3 - 2.5
Sample Compartment Size Accepts standard 10 mm pathlength glass or disposable cuvettes. Test tube and heated cell holders are available accessories.
Dimensions (WxHxD) 325 x 133 x 225 mm
Weight 1.6 kg
Voltage 90/265 VAC
Frequency 50/60 Hz
Current 15 VA
Warranty 2 years
Regulatory CE
Catalog Number Description
80-3007-12 Test tube cell holder (spring loaded for 10-18mm diameter test tubes)
80-3007-13 Heated cell holder
R-Recommended      S-Suitable      NS-Not suitable GeneQuant 100 Novaspec Pro Novaspec III+ Ultrospec 10
DNA/RNA quantitation R NS NS NS
PCR/sequencing/hybridization Primer quantitation and design S NS NS NS
Protein determination (Bradford, Biuret, BCA etc) R R NS NS
Cell culture OD600 R R NS R
Enzyme activity kinetics at 340 nm NS R R NS
Fluorescent probe quantitation, cDNA probes for microarrays, PCR probes S NS NS NS
European  pharmacopeia  compatibility NS NS NS NS
Pharma method development NS NS NS NS